Myron L D-6 User Manual
Page 40

XX. TrOuBLESHOOTING CHArT
37
SYMPTOM
POSSIBLE CAuSE
Meter won’t turn ON
• Battery weak.
Inaccurate pH readings
(D-6)
• pH calibration needed.
• Contamination from residual pH buffers
or samples in sensor well.
• Calibration with expired pH buffers.
No response to pH
changes (D-6)
• pH/ORP sensor bulb is cracked.
• Electromechanical short inside the
sensor body.
pH Zero Point calibration
will not adjust down to pH
7.0 (D-6)
• pH/ORP sensor’s internal KCl solution
is depleted or contaminated.
pH readings drift or
respond slowly to changes
in buffers/samples (D-6)
or
“FAC” is displayed
repeatedly
• Temporary condition due to the pH/
ORP sensor’s memory of solution in left
in the sensor well for long periods.
• Bulb dirty or dried out.
• Reference junction clogged or coated.
Unstable Conductivity or
TDS reading
• Dirty electrodes.
Unstable Resistivity
readings
• Dirty electrodes.
• Contamination of sample due to
exposure to air resulting in samples
with resistivity >1MΩ.
Unable to calibrate
Conductivity or TDS
• Film or deposits on electrodes.
Resistivity readings much
lower than expected
• Contamination from previous sample or
from pH sensor well.
• Test Sample is contaminated by
Carbon dioxide absorbed from the air
(commonly caused by using a standing
sample).