beautypg.com

Myron L D-6 User Manual

Page 22

background image

C. Calibration Procedures

1. Conductivity or TDS Calibration

a.

Rinse conductivity cell three times with proper standard, KCl,

NaCl, or 442 (ref. Cond/TDS Standard Solutions, pg. 39).

b.

Refill conductivity cell with same standard. NaCl-14.00 is the

example used here.

c.

Press

or

.

Then press

.

The “CAL” icon will

appear on the display

(see Figure 4).

d.

Press

or

to

step the displayed value toward the standard’s value (the

value 14.32 in Figure 4 would be stepped down to 14.00) or

hold a key down to scroll rapidly through the reading.

e.

Press

once to confirm new value and end the

calibration sequence for this particular solution type. If another

solution type is also to be measured, change solution type now

and repeat this procedure.

2. Resistivity Calibration

Resistivity is the reciprocal of Conductivity. Resistivity is calibrated only

if conductivity is calibrated for the same solution type.

3. Reloading Factory Calibration (Cond or TDS)

If calibration is suspect or known to be incorrect, and no standard

solution is available, the calibration value can be replaced with the

original factory value for that solution. This “FAC” value is the same

for all Dialysate Meters, and returns you to a known state without

solution in the cell. The “FAC” internal electronics calibration (which

bypasses the electrodes and cell) is not intended to replace calibration

with conductivity standard solutions. If another solution type requires

resetting, change solution type and repeat this procedure.

CAuTION:

DO NOT reset the calibration values or any range of your

Digital Dialysate Meter™ to factory settings unless specifically

directed to by a member of the technical staff of your facility.

Factory settings are not a substitute for calibration with reference

solutions.

ALWAYS perform dialysate checks with an instrument that has

been calibrated with reference buffers/solutions.

Figure 4

°C

NaCl

COND

CAL

µS

19