beautypg.com

Myron L D-6 User Manual

Page 20

background image

DO NOT return used samples to the storage bottle.

• This includes any solution poured into secondary containers for

use with syringes and pipettes.

DO NOT use any solution that is past its expiration date.

B. Calibration Intervals

There is no simple answer as to how often you should calibrate an

instrument. Myron L Digital Dialysate Meters™ are designed to not

require frequent recalibration. The most common sources of error were

eliminated in the design, and there are no mechanical adjustments.

Still, to ensure specified accuracy, any instrument must be checked

against chemical standards occasionally.

NOTE: If you are using the Dialysate Meter to check dialysate

concentrations, you should check the calibration of the pH and

Conductivity functions at the beginning and end of each work day

or more often if required by your clinic’s internal procedures. Then

recalibrate if indicated by the table on page 18.

On the average, we expect calibration need only be checked monthly

for the Conductivity, RES, or TDS functions. The pH (D-6) function

should be checked daily to ensure the pH/ORP sensor is not broken

or damaged. Measuring some solutions will require more frequent

intervals (ref. Suggested Calibration Intervals, pg. 23).

C. Rules for Calibration of the Dialysate Meter

1. Calibration Steps

a. Starting Calibration

Calibration is begun by pressing

while measuring Conductivity,

TDS or pH. Measuring continues, but the “CAL” icon is on, indicating

calibration is now changeable.

The reading is changed with the

and

keys to match the

known value. The calibration for each of the 3 solution types may be

performed in either conductivity or TDS mode.

Depending on what is being calibrated, there may be 1, 2 or 3 steps to

the calibration procedures.

Unless otherwise instructed by your clinic’s internal procedures, use

the following table

to determine when calibration is indicated for specific

functions.

17

Proper Handling of Reference Solutions continued from previous page