beautypg.com

Myron L D-6 User Manual

Page 25

background image

3.

Rinse pH/ORP sensor well and conductivity cell 3 times with

acid or base buffer solution.

4.

Refill pH/ORP sensor well and conductivity cell again with

same buffer solution.

5.

Press

or

until display agrees with buffer value.

6.

Press

to accept 2nd point of calibration. Now the

display indicates the next type of buffer to be used.

Single point Gain Calibration is complete. Continue for the 3rd point of

Calibration (2nd Gain).

DO NOT press any measurement keys as this will exit calibration. Exiting

causes the value accepted for the buffer to be used for both acid and

base measurements.

To continue with 3rd point calibration, use basic buffer if acidic buffer

was used in the 2nd point, or vice-versa. Again, match the display to the

known buffer value as in step 2 and continue with the following steps:
7.

Repeat steps 3 through 6 using opposite buffer solution.

8.

Press

to accept 3rd point of calibration, which

completes the Calibration procedure. Fill pH/ORP sensor

well and conductivity cell with Myron L Storage Solution and

replace protective cap.

6. ORP Calibration (D-6)

ORP electrodes rarely give false readings without problems in the

reference electrode. For this reason, and because calibration solutions

for ORP are highly reactive and potentially hazardous, your Dialysate

Meter has an electronic ORP calibration. This causes the zero point on

the reference electrode to be set whenever pH 7 calibration is done.

7. Temperature Calibration

Temperature calibration is not necessary. The Dialysate Meter temperature

sensing circuit uses a high quality sensor that does not require calibration.

Figure 6

BUFFER

pH

CAL

Figure 7

pH

BUFFER

CAL

22