Myron L EP-10 User Manual
4&5 range conductivity meters, Description
description
This manual describes your Myron L EP series meter,
tells you how to use it, and how to keep it working
accurately for many years.
Myron L EP & EP10 Meters are compact, multi-range
instruments which operate on the principle of electrical
conductivity. They each contain a built-in cell cup and will
quickly determine the conductivity of almost any solution.
Both have dials calibrated in micromhos (µmhos)
equivalent to the metric microsiemens (µS). Model EP
features a second dial scale calibrated in
megohms/resistivity for testing ultrapure water.
Both are 3.4” x 4.5” x 4.0” (86 x 114 x 102mm) and
weigh less than one pound (0.45 kg). EP series meters are
completely self-contained. The built-in cell is automatically
temperature compensated. They are powered by one 9
volt battery good for at least 2000 tests, or one year shelf
life.
model ranges
ep
0-0.5, 0-5, 0-50, 0-500, 0-5,000 micromhos (µmhos)
2-30 Megohms
ep10 0-10, 0-100, 0-1,000, 0-10,000 micromhos (µmhos)
accuracy: (Both Models) ±2% of full scale
repeatability: (Both Models) ±1%
temperature compensation to 25ºc:
ep
32-120ºF (0-49ºC)
ep10
50-160ºF (10-71ºC)
By using an RE-10 Range Extender (see
aCCESSoRiES), the maximum range of each meter may
be increased ten times.
The pictures in this manual show the major operating
parts of your Myron L EP series meter. Handle your meter
and identify these parts to become familiar with it.
Battery
Indicator
Range Switch
Cell Cup
Button
Upper Electrode
Meter Zero
Adjust
Pointer
Model EP10 shown
4&5 range conductivity meters
oPERaTing inSTRuCTionS
User’s Manual for Models EP
EP-10