XVI. CELL CHECK
The cell check verifies the cleanliness of the conductivity/TDS/resistivity

sensor. In normal use the cell may become dirty or coated and require

cleaning. If the display is showing “.00” when the cell cup is dry, the sensor is

probably clean. However, when testing high purity water in resistivity (RES)

mode improved accuracy may be desired. No matter what a manufacturer

claims, a sensor can and will become contaminated or coated and will

therefore require cleaning. A true 4-wire sensor, as in the Dialysate Meter,

helps to mitigate contamination, but NO SENSOR IS 100% IMMUNE.

1.

Press

.

2.

Press

to display the

stored memory records.

3.

Press

repeatedly

until you pass the “FAC SEL”

location. The display will show

“CELL ch” (see Figure 20).

4.

Press

to test.

If cell is clean, “Good” will

momentarily be displayed.

(See Figure 21.) If cell is dirty,

“CELL cLn” will be displayed

(see Figure 22). (Ref.

Cleaning Sensors, pg. 34.)

XVII. AuTO OFF

Auto off allows the user to adjust the

time the instrument is ON (up to 75

seconds) after each press of a key.

Default time is 15 seconds in

measurement mode and 60

seconds in “CAL” (calibration) mode.