beautypg.com

Myron L D-6 User Manual

Page 11

background image

IV.

INITIAL CALIBrATION and FACTOrY SETTINGS

CAuTION:

The factory settings are not intended to replace calibration

with reference solution/buffers.

Factory settings do not take into account the current condition of

the actual sensor installed in your instrument or any environmental

conditions that might exist in your facility.

DO NOT use a Digital Dialysate Meter™ to check the pH or

conductivity of dialysate using factory calibration settings.

ALWAYS perform an initial calibration of a Digital Dialysate

Meter™ with reference solutions/buffers prior to putting it into use

for the first time.

The following table lists initial calibration procedures and their location

in this manual.

PARAMETER PROCEDURE REFERENCE SOLUTION
General instructions (see Calibration, pg. 16)
Conductivity

Gain

14.00 mS NaCl

1

(see Conductivity or

TDS Calibration, pg. 19)

Resistivity

Electronically calibrated with Conductivity Gain

TDS

Gain

442™

1

(see Conductivity or TDS

Calibration, pg. 19)
3000 PPM

1

(see Conductivity or TDS

Calibration, pg. 19)

pH

Zero Point

7.0 pH (see pH Zero Calibration

(D-6), pg. 20)

Gain – Acid

4.0 pH (see pH Gain Calibration

(D-6), pg. 21)

Gain – Base

10.0 pH (see pH Gain Calibration

(D-6), pg. 21)

ORP

Electronically calibrated with pH Zero Point

Temperature Does not require calibration

1

Make sure the correct solution type is selected and that it matches the type of

reference solution being used (see Solution Selection, pg. 15)

Avoid letting solvents come into contact with the instrument’s sensors,

case or window areas. Shock damage from a fall may cause the

instrument to fail.

WArNING:

You MuST verify the function and calibration of any Dialysate

Meter that has been dropped by performing calibration checks (see

Checking pH/ORP Sensor Performance, pg. 34) of all measurement

functions prior to putting it back into use.

8