NaCl conversion factor.

DO NOT change the calibration solution type unless specifically

directed to do so by a member of the technical staff of your facility.

NOTE: Check display to see if solution displayed (KCl, NaCl or 442) is

already the type desired. If not:

1.

Press

,

or

to select the parameter whose

solution type you wish to change.

2.

Press and hold

key

for 3 seconds to make

“SEL” appear (see Figure

3). For demonstration

purposes, all 3 solution

types are shown

simultaneously.

3.

Use the

or

key to select type of solution desired

The selected solution type will be displayed: KCl, NaCl or 442.

4.

Press

to accept new solution type.

X.

CALIBrATION

A. Proper Handling of Reference Solutions

ALWAYS perform calibrations and calibration checks with standard

reference solutions that are NIST traceable.

• Myron L Company reference solutions are highly recommended

and are the same solutions used in the manufacture and factory

calibration of your Digital Dialysate Meter™.

To ensure standard solution and calibration/verification accuracy

• Store solutions in a cool, dark place.

• Preferred storage temperature is 25°C/77°F, the ideal calibration

temperature.

• Keep solutions tightly capped to avoid evaporation.

• When using a solution, pour it directly from the bottle.

• If using a syringe or pipette, pour the solution into another

container and fill the syringe or pipette from there.

DO NOT use the pipette or syringe to draw solution directly from

the bottle.

• Use the solution immediately after dispensing to prevent changes in

concentration due to evaporation.

• Recap the bottle immediately after dispensing.

• Keep the bottle capped when not in use.

Figure 3

KCl

442

NaCl

16