beautypg.com

Myron L D-6 User Manual

Page 38

background image

• If the pH reading is between 8.0 pH and 6.0 pH, calibrating the

sensor (see pH Calibration, pg. 20) will correct the offset.

7.

Press

to return the instrument to measurement mode

without changing the calibration of the instrument.

3. pH/ORP

CAuTION:

ONLY authorized members of your facility’s maintenance

staff should attempt the following procedures.

The unique pH/ORP sensor in your Dialysate Meter is a nonrefillable

combination type that features a porous liquid junction.

It should not be

allowed to dry out. However, if this occurs, the sensor may sometimes

be rejuvenated by first cleaning the sensor well with isopropyl alcohol

or a liquid spray cleaner, such as Windex™ or Fantastic™, and rinsing

well. Do not scrub or wipe the pH/ORP sensor.

Then use one of the following methods:
1.

Pour a HOT salt solution ~60°C/140°F — preferably potassium

chloride (KCI) solution (Myron L pH/ORP Sensor Storage

Solution) but HOT tap water with table salt (NaCl) will work

fine — in the pH/ORP sensor well and allow to cool. Retest.

or
2.

Pour DI water in the pH/ORP sensor well and allow to stand for

no more than 4 hours (longer can deplete the reference

solution and damage the glass bulb). Retest.

If neither method is successful, the sensor must be replaced.

Drifting can be caused by a film on the pH sensor bulb and/or

reference (see Figure 31). Use isopropyl alcohol or spray a liquid

cleaner such as Windex™ or Fantastic™ into the sensor well to clean

it. The sensor bulb is very thin and delicate. Do not scrub or wipe the

pH/ORP sensor.

pH/ORP SENSOR

Top View

ORP

Electrode

pH Glass

Electrode

Sensor

Body

Reference

Junction

under Glass

pH Bulb

Figure 31

35