Myron L D-6 User Manual
Page 27

B. Calibration Tracking Records for General Use (Not Dialysate)
To minimize your calibration effort, keep records. If adjustments you
are making are minimal for your application, you can check less often.
Changes in conductivity calibration should be recorded in percent.
Changes in pH calibration (D-6) are best recorded in pH units.
Calibration is purposely limited in the Dialysate Meter to ±10% for the
conductivity cell, as any change beyond that indicates damage, not
drift. Likewise, calibration changes are limited to ±1 pH unit, as any
change beyond that indicates the end of the sensor’s lifetime and
replacement is recommended.
WArNING:
For Dialysate measurements, it is critical that you verify that
the pH sensor is working at the beginning and end of each workday
or more often if required by your clinic’s internal procedures.
C. Conductivity, RES, TDS Practices to Maintain Calibration
Note: Very active chemicals may discolor the conductivity electrodes
(though this is not likely in clinical use).
• This does not affect the accuracy of conductivity measurements.
1.
Clean oily films or organic material from the cell electrodes by
pouring isopropyl alcohol or squirting a foaming non-abrasive
cleaner into the cell and rinsing (see Cleaning Sensors, pg. 34).
• DO NOT scrub inside the cell.
2.
Calibrate with solutions close to the measurements you make.
Readings are compensated for temperature based on solution
type. If you choose a solution type that does not closely match
the characteristics of the solution you are trying to measure, your
records of calibration changes will reflect temperature changes
more than changes in the instrument’s accuracy. For example,
if you choose to measure tap water solute concentration in KCl
mode, but you calibrate with 442 because it is handy, the further
away from 25ºC the tap water sample is, the more error there
will be in the temperature-compensated reading.
Note: Dialysate
conductivity measurements MuST be set to NaCl.
3.
Rinse out the cell with pure water after taking measurements.
• Allowing slow dissolving crystals to form in the cell
contaminates future samples.
4.
For maximum accuracy, keep the pH sensor cap on tight so that
no fluid in the pH sensor well washes into the conductivity cell.
D. pH and ORP Practices to Maintain Calibration (D-6)
1.
ALWAYS keep the sensor wet with Myron L pH Sensor
Storage Solution when not in use. If the pH/ORP sensor is
allowed to dry out, its accuracy will be significantly
compromised.
24