WArNING:

The case gasket MuST be flat and inserted all the way into

the groove on the top case with NO kinks or twists.

• An improperly seated case gasket will allow moisture to leak into

case that will damage the instrument.

Check the four screws.

• Each screw MuST have a small black O-ring around its shaft just

under the head.

• Take care not to lose or damage the O-rings.

Dry Instrument THOrOuGHLY. Remove the four (4) bottom screws.

Open instrument carefully. Carefully detach battery from circuit board.

Replace with 9 volt alkaline battery. Replace bottom, ensuring the

sealing gasket is installed in the groove of the top half of case.
Tighten each screw until snug in the order shown in Figure 30. Then

tighten each screw another ½ turn in the order shown in Figure 30.
NOTE: Though all calibration settings are protected

even during power loss or battery replacement, loss

of time and date may occur if battery is removed

for longer than 3 minutes (180 seconds).

C. Preserving pH/ORP Sensor Function (D-6)

It is important to read and follow the instructions found

in pH and ORP Practices to Maintain Calibration,

pg. 24. In addition to these precautions:

DO NOT leave high pH (alkaline) solutions

in contact with the sensor for long periods of

time. This will damage it.

• Samples containing chlorine, sulfur,

or ammonia can poison the sensor.

• If it is necessary to measure the pH of any such solutions,

thoroughly rinse the pH/ORP sensor well and conductivity

cell with clean water immediately after taking the

measurement.

DO NOT use a Digital Dialysate Meter™ to measure samples

containing substances that reduce silver, such as cyanide.

• These substances will damage the pH/ORP sensor.

D. pH/ORP Sensor Replacement (D-6)

WArNING:

Only authorized members of your facility’s maintenance staff

should attempt pH/ORP sensor replacement. Complete installation

instructions are provided with each replacement sensor.

WArNING:

After replacing the pH/ORP sensor of your D-6 Dialysate

Meter and PrIOr to putting it into use, ALWAYS perform a complete

calibration of the following measurement functions with reference

solutions/buffers:

• pH

Figure 30

33