beautypg.com

Myron L D-6 User Manual

Page 17

background image

4.

Press

.

5.

Take reading.

NOTE: When ppm free chlorine units are selected (ref. VIII. ORP UNIT

SELECTION), annunciators alert you when the concentration is outside

the specified measurement range. “Or” (over range) will display when

the concentration is over the range limit (> 9.9 ppm). “ur” (under range)

will display when the concentration is below the range limit (< 0.2 ppm).

WArNING:

When using a D-6 to verify the complete removal of residual

oxidizing germicides, such as chlorine, ORP mV reading should be

as low as possible. Generally, an ORP reading of < 275 mV is a good

indicator that there is very little residual oxidizer remaining. However,

it is recommended that a target reading matching or close to the ORP

value measured at the output of your RO system be sought.

6.

IMPOrTANT: After use, fill pH/ORP sensor well with Myron L pH

Sensor Storage Solution and replace protective cap. If Myron L

pH Sensor Storage Solution is unavailable, you can use a strong

KCl solution (preferred), a saturated solution of NaCl table salt and

tap water (next best alternative), or a pH 4 buffer (least desirable

alternative). NEVEr use distilled water or plain tap water as this

will quickly deplete the sensor (ref. Cleaning Sensors, 3. pH/ORP,

pg. 35).

Do not allow pH/ORP sensor to dry out.

VIII.

OrP uNIT SELECTION (D-6) mV Or PPM FrEE CHLOrINE

1.

Press

.

2.

Press and hold

for approximately 3 seconds.

The current preference for

ORP units of measure is

displayed. Factory setting for

this preference is mV (see

Figure 1).

4.

Press the

or

keys

to toggle between mV

and ppm free chlorine

measurement format. The

setting chosen is displayed.

(ppm shown as example in

Figure 2).

mV

Figure 1

14

PPM

Figure 2