Myron L D-6 User Manual

Page 23

background image

a.

Press

or

.

b.

Press

.

c.

Press

key until “FAC” appears and release.

d.

Press

to accept the factory calibration setting.

4. pH Calibration (D-6)

ALWAYS calibrate a Digital Dialysate Meter™ pH function at Zero

Point with a pH 7.0 buffer solution BEFOrE performing either the

Acid or Base gain calibration.

• The Acid/Base gain calibrations may be done in either order.

Read Proper Handling of Reference Solutions, pg. 16, before

attempting pH calibration.
Calibrating the pH function of a Digital Dialysate Meter™ is a three-

step process. Zero Point calibration is always done first. Steps 2 and 3

can be performed in either order.

STEP 1: Zero Point Calibration (7.0 pH buffer)

STEP 2: Gain Calibration – Acid (4.0 pH buffer)

STEP 3: Gain Calibration – Base (10.0 pH buffer)

a. pH Zero Calibration (D-6)

1.

Rinse pH/ORP sensor well and conductivity cell 3 times with 7

buffer solution.

2.

Refill both sensor wells with 7 buffer solution.

3.

Press

to verify the pH calibration. If the display shows

7.00, skip the pH Zero Calibration and proceed to section b.

pH Gain Calibration, pg. 21.

4.

Press

to enter

calibration mode. The

“CAL”, “BUFFER” and “7”

annunciators will appear

(see Figure 5).

Displayed value will be the

uncalibrated sensor.

Figure 5

BUFFER

pH

CAL

20