beautypg.com

Figure v.a.1, Primary component identification – Myron L Aquaswitch II User Manual

Page 3

background image

QUICK SET POINT ADJUSTMENT -

S e e S e c t i o n I V . B . 2 .

The set point setting is based upon the user's particular water
purity specifications or requirements.

1. While pressing the "SET POINT" switch, turn the Set

Point #1 adjustment screw (see figure V.A.1) until the
desired set point value is indicated on the display.

HYSTERESIS (DEAD BAND) ADJUSTMENT -

S e e S e c t i o n I V . B . 3 .

PRIMARY COMPONENT IDENTIFICATION -

Section V.A

.

Review the figure below to familiarize yourself with the different
circuit boards and component locations. The diagram has the
second alarm/control module option installed.

“PURGE CYCLE” CALIBRATION PROCEDURE

Section V.B.

“ALARM RESET” MODE CHANGE PROCEDURE

Section V.C.

QUICK CALIBRATION - Section V.D

.

WARNING: When performing calibration procedures,
the technician must take extreme care to avoid
contacting the circuitry other than the CALibration
control. Failure to do so could result in damage to
the equipment, property and/or personal injury.

The following assumes the front panel has been
removed and the power is ON.

ELECTRONIC CALIBRATION (CIRCUIT ONLY) -

S e e S e c t i o n V . D . 1 .

Full Scale Calibration V.D.1.a.

1.

Press and hold the Full Scale Test switch. The display
should indicate Full Scale for the particular range
selected, i.e. 0-20 M

should indicate 20. If not, set to

Full Scale with the CALibration control.

2.

Turn power O F F .

3.

Re-install front panel as described in “REASSEMBLY”.

4.

To operate, turn power ON.

10VDC Calibration - See Section V.D.1.b.

USING STANDARD SOLUTIONS - See Section V.D.2.
The BEST method of verifying and recalibrating your
conductivity/TDS

AQUASWITCH II

is with NIST traceable

Standard Solution (available from the Myron L Company).
Because it includes the sensor, the entire system is recalibrated.
NOTE: Since standard solution calibrations are NOT practicable
with resistivity models, another means of verification or
calibration of resistivity models is to use the transfer standard
method, using a hand-held or portable instrument capable of
resistivity measurements, such as the Myron L Ultrameter™. See
section V.C.4 for description.

SENSOR SUBSTITUTE CALIBRATION -

S e e S e c t i o n V . D . 3 .

TRANSFER STANDARD METHOD - See Section V.D.4.

21 may 03

Figure V.A.1

AQUASWITCH II

MAIN MONITOR/CONTROLLER BOARD

E
N
C
L
O
S
U
R
E

PRIMARY COMPONENT IDENTIFICATION

AQUASWITCH II

AQUASWITCH I

AQUASWITCH

LOGIC BOARD

F
R
O
N
T

P
A
N
E
L

LOW

DISPLAY
HARNESS

DISPLAY

DISPLAY CONNECTOR

RED

GREEN

SET POINT

SWITCH

TO MONITOR/
CONTROLLER
CIRCUIT BOARD
DISPLAY
CONNECTOR

AQUASWITCH II

DIGITAL DISPLAY

CABLE
CONNECTS TO
POWER BOARD

PURGE CYCLE
CONTROL

AQUASWITCH

POWER BOARD

TRANSFORMER

FUSE*

115/230
SWITCH

AQUASWITCH™

POWER BOARD

NC

NO

COM

GRD

C

AC

NO

COM

NC

FUSE
1/8 Amp
AGC 1/8

BK B

BK A

BLD

COM

LOGIC
BOARD
CONNECTION

RELAY

115/230
JUMPERS

OR

ALARM OUTPUT
CONNECTORS

*115/230VAC
FUSE
100mA
(T.10A) 5X20
Slow Blow/
Time Delay

*24 VAC/24 VDC
FUSE
250mA
(T.25A) 5X20
Slow Blow/
Time Delay

TRANSFORMER

FUSE*

1 1 5 /
230

751 756
752 757
753 758
754 759

HYS1

SP1

FS SW

DIS

INC

DEC

CAL

-11

0-20m

UP

}

3S

SPC

CALIBRATION
CONTROL

DISPLAY
HARNESS
CONNECTOR

DISPLAY
CALIBRATION
CONTROL

RANGE
MODULE

4-20
CONNECTOR

BK WT RD GN NU R- R+

REMOVE TO
INSTALL
SECOND
RELAY

PWR C GD NC NO CM

BK B

WHT

STANDARD
ALARM RESET
ENABLE JUMPER

MYRON L
COMPANY

This manual is related to the following products: