Agilent Technologies N9340A User Manual
Page 147

Error Messages
7
N9340A User’s Guide
145
147
Incorrect
alignment data
in EEPROM
Indicates invalid alignment data
occurred.
143
Final IF
overload
Indicates you need to either
increase the input attenuation or
decrease the input signal level.
138
USB device NOT
ready
Indicates the USB device is not
detected.
136
RBW limit to 30
kHz when in fast
sweep
Indicates the analyzer
automatically couple the RBW to
30 kHz when in fast sweep
mode.
135
Error 0xB
Indicates the analyzer encounters
a DLL file open failure. If
analyzer froze, try to restart the
analyzer.
134
Error 0xA
Indicates the analyzer failed at a
dynamic loading process. If
analyzer froze, try to restart the
analyzer.
132
Error 0x8
Indicates some important files
missing.
130
Meas uncal
Indicates the measurement is
uncalibrated due to fast sweeping
through a narrow RBW filter.
Check the sweep time, span, and
bandwidth settings, or use auto
coupling.
119
RF Board Changed
Indicates the RF board was
changed, the analyzer needs
re-load the alignment data.
116
Cannot
communicate with
RF
Indicates the MCU cannot find
the RF board.
112
Error 0x3, DSP
will restart
Indicates an undefined error
occurred when MCU tried to
communicate with DSP. Wait the
analyzer to reboot.
Error Code Error Message
Description