Connection rule, Connection rule -14 – Agilent Technologies B2201A User Manual
Page 74

3-14
Agilent B2200 User’s Guide, Edition 2
Front Panel Operation
Switch Control Functions
Connection Rule
Connection rule is used to specify if an input or output port can have multiple 
connections. For each module installed in the mainframe (for auto configuration 
mode, installed modules are treated as one module), you can specify one of the 
following connection rules:
Single:
Each input port can be connected to only one output port, and 
each output port can be connected to only one input port. So, 
existing connection to a port will be disconnected when a new 
connection is made.
If normal configuration mode is set to the Agilent B2200 with 
multiple modules, an input/output path is effective for each 
module. For example, if normal configuration mode is set to the 
Agilent B2200 with two modules, and single rule is set to both 
modules, the following connection is possible.
Free:
Each input port can be connected to multiple output ports, and 
each output port can be connected to multiple input ports. 
CAUTION
If the Free connection rule has been specified, ensure multiple input ports are not 
connected to the same output port. Such configurations can cause damage.
Module 1 Output
1
12
2
1
Close
Open
14
3
2
Module 2 Output
1
2
12
3
3
Input