System:cdescription, System:cdescription? -45 – Agilent Technologies B2201A User Manual

Page 171

background image

Agilent B2200 User’s Guide, Edition 2

5-45

SCPI Command Reference

:SYSTem:CDEScription?

:SYSTem:CDEScription?

This query command returns a description of the specified card.

Syntax

:SYSTem:CDEScription? card_number

Query response

"card_description" <^END>

If the designated card is not installed or initialize failed, this command returns
"No Card".

For B2200A:

In the Auto Config mode, "card_description" response is one of following:

• "B2210A 14x12 Femto Leakage Switch Module"
• "B2210A 14x24 Femto Leakage Switch Module"
• "B2210A 14x36 Femto Leakage Switch Module"
• "B2210A 14x48 Femto Leakage Switch Module"

In the Normal Config mode:

"B2210A 14x12 Femto Leakage Switch Module"

For B2201A:

In the Auto Config mode, "card_description" response is one of following:

• "B2211A 14x12 Low Leakage Switch Module"
• "B2211A 14x24 Low Leakage Switch Module"
• "B2211A 14x36 Low Leakage Switch Module"
• "B2211A 14x48 Low Leakage Switch Module"

In the Normal Config mode:

"B2211A 14x12 Low Leakage Switch Module"

Example

OUTPUT @Agb2200;":SYST:CDES? 1"

ENTER @Agb2200;A$

Parameter Explanation

card_number

For Auto Config mode: 0
For Normal Config mode: 1, 2, 3, or 4