Agilent Technologies B2201A User Manual

Page 123

background image

Agilent B2200 User’s Guide, Edition 2

4-25

Programming

Capacitance Compensation

Table 4-10

Compensation Coefficients and Modifications

2. Measure the R, L, C values of the C2x or C3x path by using the Agilent 4284A.

See “To obtain compensation coefficients” on page 4-26.

After the measurements, calculate the per meter value of the R, L, C, and record
them into the following table.

3. Open the template file selected at step 1 by using a text editor. Exchange the R,

L, C values of C2x/C3x with the values recorded at step 2. And save the file as
your compensation data file (e.g. C:\temp\my_env_1.txt).

Do not change any other lines. Also do not change the value for the coefficients
that should not be modified.

Compensation

coefficients

Modifications of data file

C2H

C2L

For the Agilent B2220A probe card interface, do not modify
the lines.

For the connector plate, change the R, L, C values in the lines.
The value must be changed to the R, L, C values of the C2x
path (triaxial cable with connector plate) shown in Figure 4-6.

C3H

C3L

Change the R, L, C values in the lines. The value must be
changed to the R, L, C values of the C3x path.

For the Agilent B2220A probe card interface, probe card will
be used for the C3x path.

For the connector plate, coaxial cable with positioner will be
used for the C3x path.

Compensation

coefficients

Explanation

R (

)

L (H)

C (F)

C2H

C2L

C3H

C3L