Agilent Technologies Agilent 4396B User Manual
Page 76
![background image](/manuals/44282/76/background.png)
Analyzer
Settings
Press
4
Preset
5 .
Then
set
the
analyzer's
controls
as
follows:
Desired
Settings
K
ey
Strokes
A
ctive
Channel
Block
Select
channel
1
4
Chan
1
5
(default)
Measurement
Block
Select
Spectrum
Analyzer
Press
4
Meas
5
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
ANALYZER
TYPE
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
NETWORK
ANALYZER
Select
S
21
(or
B/R)
measurement
4
Meas
5
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
Trans:FWD
S21
[B/R]
Select
LOG
MA
G
format
4
F
ormat
5
FFFFFFFFFFFFFFFFFFF
LOG
MAG
(default)
Sweep
Block
Center
frequency
250
MHz
Press
4
Center
5
250
4
M/
5
Span
frequency
200
MHz
Press
4
Span
5
200
4
M/
5
P
erforming
Calibration
P
erform
a
frequency
response
calibration
for
this
measurement
as
follows:
1.
Press
4
Cal
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CALIBRATE
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNN
RESPONSE
.
2.
Connect
a
THRU
calibration
standard
between
the
measurement
cables
in
place
of
the
DUT
.
3.
Press
NNNNNNNNNNNNNN
THRU
to
perform
a
frequency
response
calibration
data
measurement.
4.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE:RESPONSE
.
(
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION
on
OFF
is
automatically
set
to
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION
ON
off
.)
Network
Measurement
1.
Press
4
Scale
Ref
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTO
SCALE
to
t
scale
to
the
trace
.
2.
Press
4
Search
5
NNNNNNNNNNNNNN
PEAK
to
put
the
marker
on
an
unknown
peak
of
the
trace
.
The
transmission
measurement
result
is
displayed
as
shown
in
Figure
4-9.
Figure
4-9.
Transmission
Measurement
Result
(In uenced
by
Leakage)
4-8
Spectrum
Measurement
Examples