Agilent Technologies Agilent 4396B User Manual
Page 110
With
a
bit
in
the
I/O
port
on
the
rear
panel.
With
the
GPIB
commands
OUTPLIMF? ,
OUTPLIMIL?
,
and
OUTPLIMM? .
Separated
Limit
Lines
Figure
5-21
shows
separated
limit
lines
and
an
editing
table
example
.
This
example
can
be
used
for
lter
testing
that
only
requires
insertion
loss
limits
.
Dummy
limit
values
(+5000
for
upper
and
05000
for
lower
,
for
example)
should
be
entered
for
the
no
limit
areas
.
Figure
5-21.
Separated
Limit
Lines
Limits
are
only
checked
at
each
of
the
actual
measured
data
points
.
It
is
possible
for
a
device
to
be
out
of
specication
without
a
limit
test
failure
indication
if
you
do
not
select
sucient
sweep
parameter
points
within
a
segment.
By
combining
the
limit
test
and
the
list
sweep
function,
high
throughput
limit
line/testing
can
be
performed
for
go/no-go
testing.
F
or
detailed
information
on
the
list
sweep
,
see
\List
Sweep
"
in
this
chapter
.
Limit
line
information
is
lost
if
you
press
4
Preset
5
or
turn
o
the
power
.
However
,
the
4
Save
5
keys
can
save
the
limit
line
information
along
with
all
other
current
instrument
settings
when
the
limit
lines
are
on.
See
the
T
ask
R
eference
manual
for
details
.
5-24
Network
Measurement
Examples