4. THERMOCOUPLE INPUTS:

Input Impedance: 20 MΩ

Max. Continuous Overvoltage: 30 VDC

Failed Sensor Indication: 

*After 20 min. warm-up. Accuracy is specified in two ways: Accuracy at

23°C and 15 to 75% RH environment; and Accuracy over a 0 to 50 °C and 0 to

85% RH (non condensing) environment. Accuracy specified over the 0 to 50

°C operating range includes meter tempco and cold junction tracking effects.

The specification includes the A/D conversion errors, linearization

conformity, and thermocouple cold junction compensation. Total system

accuracy is the sum of meter and probe errors. Accuracy may be improved by

field calibrating the meter readout at the temperature of interest.

5. RTD INPUTS:

Type: 2, 3 or 4 wire

Excitation Current:

100 ohm range: 165 µA; 10 ohm range: 2.5 mA

Lead Resistance:

100 ohm range: 10 Ω/lead max.; 10 ohm range: 3 Ω/lead max.

Balanced Lead Resistance: Automatically compensated up to max per lead

Unbalanced Lead Resistance: Uncompensated

Max. Continuous Overvoltage: 30 VDC

Failed Sensor Indication:  or 

*After 20 min. warm-up. Accuracy is specified in two ways: Accuracy at

23°C and 15 to 75% RH environment; and Accuracy over a 0 to 50°C and 0 to

85% RH (non condensing) environment. Accuracy specified over the 0 to 50°C

operating range includes meter tempco effects.

The specification includes the A/D conversion errors and linearization

conformity. Total system accuracy is the sum of meter and probe errors.

Accuracy may be improved by field calibrating the meter readout at the

temperature of interest.

6. USER INPUT: Programmable input

Software selectable for active logic state: active low, pull-up (24.7 KΩ to +5

VDC) or active high, pull-down resistor (20 KΩ).

Trigger levels: V

IL

= 1.0 V max; V

IH

= 2.4 V min; V

MAX

= 28 VDC

Response Time: 10 msec typ.; 50 msec debounce (activation and release)

7. MEMORY: Nonvolatile E

2

PROM retains all programming parameters and

max/min values when power is removed.

8. OUTPUTS:

Type: Dual Form C contacts

Isolation to Sensor & User Input Commons: 1400 Vrms for 1 min.

Working Voltage: 150 Vrms

Contact Rating: 5 amps @ 120/240 VAC or 28 VDC (resistive load), 1/8

H.P. @ 120 VAC (inductive load)

Life Expectancy: 100 K cycles min. at full load rating. External RC snubber

extends relay life for operation with inductive loads.

Response Time: Turn On or Off: 4 msec max.

9. ENVIRONMENTAL CONDITIONS:

Operating temperature: 0 to 50 °C

Storage temperature: -40 to 70 °C

Operating and storage humidity: 0 to 85% max. RH (non-condensing)

Vibration to IEC 68-2-6: Operational 5 to 150 Hz, 2 g.

Shock to IEC 68-2-27: Operational 30 g (10 g relay).

Altitude: Up to 2,000 meters

10. CONNECTIONS: High compression cage-clamp terminal block

Wire Strip Length: 0.3" (7.5 mm)

Wire Gage: 30-14 AWG copper wire

Torque: 4.5 inch-lbs (0.51 N-m) max.

11. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 outdoor use. IP20

Touch safe. Installation Category II, Pollution Degree 2. One piece bezel/

case. Flame resistant. Synthetic rubber keypad. Panel gasket and mounting

clip included.

12. CERTIFICATIONS AND COMPLIANCES:

CE Approved

EN 61326-1 Immunity to Industrial Locations

Emission CISPR 11 Class A

IEC/EN 61010-1

RoHS Compliant

Type 4X Outdoor Enclosure rating (Face only)

IP65 Enclosure rating (Face only)

IP20 Enclosure rating (Rear of unit)

For Model No. PAXLT0U0 Only: UL Listed: File #E137808

Refer to EMC Installation Guidelines section of the bulletin for additional

information.

13. WEIGHT: 10.4 oz. (295 g)

TC TYPE

RANGE

ACCURACY

@ 23°C

±

°C *

WIRE COLOR

5.8

4.3

15.0

ANSI

BS 1843

T

-200 to 400°C

-328 to 752°F

2.3

(+) blue

(-) red

(+) white

(-) blue

J

-200 to 760°C

-328 to 1400°F

1.9

(+) white

(-) red

(+) yellow

(-) blue

R

-50 to 1768°C

-58 to 3214°F

4.5

no

standard

(+) white

(-) blue

S

-50 to 1768°C

-58 to 3214°F

4.5

no

standard

(+) white

(-) blue

B

200 to 1820°C

392 to 3308°F

9.1<540°C

4.5>540°C

no

standard

no

standard

E

-200 to 871°C

-328 to 1600°F

2.7

(+) purple

(-) red

(+) brown

(-) blue

N

-200 to 1300°C

-328 to 2372°F

2.8

(+) orange

(-) red

(+) orange

(-) blue

K

-200 to 1372°C

-328 to 2502°F

2.3

(+) yellow

(-) red

(+) brown

(-) blue

mV

C

(W5/W26)

-10.00 to 65.00

0 to 2315°C

32 to 4199°F

0.02 mV

1.9

no

standard

no

standard

no

standard

no

standard

15.0

42.6<540°C

15.0>540°C

4.9

8.1

5.8

0.08 mV

6.1

ACCURACY

@ 0 to 50°C

±

°C *

STANDARD

ACCURACY*

@0 to 50°C

ACCURACY*

@ 23°C

RANGE

no official

standard

0.9°C

0.4°C

-100 to 260°C

10 ohm Copper

alpha = .00427

RTD TYPE

no official

standard

0.5°C

0.2°C

-80 to 260°C

120 ohm Nickel

alpha = .00672

no official

standard

1.6°C

0.4°C

-200 to 850°C

100 ohm Pt

alpha = .00392

IEC 751

1.6°C

0.4°C

-200 to 850°C

100 ohm Pt