Vaisala WXT510 User Manual
Page 87

Chapter 7 __________________________________________________ Getting the Data Messages
VAISALA _______________________________________________________________________ 85
Example of the XDR Query (all parameters of each sensor enabled
and NMEA wind formatter set to T):
$--WIQ,XDR*2D
Example of the response when all the parameters of each sensor are
enabled (NMEA wind formatter set to T):
Wind sensor data
$WIXDR,A,302,D,0,A,320,D,1,A,330,D,2,S,0.1,M,0,S,0.2,M,1,S,0.2,
M,2*57
P, T, and RH data
$WIXDR,C,23.3,C,0,C,24.0,C,1,H,50.1,P,0,P,1009.5,H,
0*75
Precipitation data
$WIXDR,V,0.02,M,0,Z,30,s,0,R,2.7,M,0,V,0.0,M,1,Z,0,s,1,R,0.0,M,1,
R,6.3,M,2,R,0.0,M,3*51
Supervisor data
$WIXDR,C,25.5,C,2,U,10.6,N,0,U,10.9,V,1,U,3.360,V,2*71
Internal temperature
1
Relative humidity
0
Rain accumulation
0
Rain duration
0
Rain current intensity
0
Hail accumulation
1
Hail duration
1
Hail current intensity
1
Rain peak intensity
2
Hail peak intensity
3
Heating temperature
2
Supply voltage
0
Heating voltage
1
3.5 V reference voltage
2
Table 6
Transducer IDs of the Measurement Parameters
(Continued)
Measurement
Transducer ID