Setting fields – Vaisala WXT510 User Manual
Page 102

User’s Guide ______________________________________________________________________
100 _________________________________________________________________ M210470EN-D
Setting Fields
[R]
=
Parameter selection: This field consists of 16 bits
defining the PTU parameters included in the data
messages. The bit value 0 disables and the bit value 1
enables the parameter.
The bits 1-8 determine the
parameters included in the
message obtained with the
following commands:
-ASCII: aR2 and ar2
-NMEA 0183: $--WIQ,XDR*hh
-SDI-12: aM2, aMC2, aC, and
aCC2
-SDI-12 continuous: aR2 and
aRC2
1st bit (most left)
Pa Air pressure
2nd bit
Ta Air temperature
3rd bit
Tp Internal temperature
1
1. Tp temperature value is used in pressure calculation, it does not express the air temperature.
4th bit
Ua Air humidity
5th bit
spare
6th bit
spare
7th bit
spare
8th bit
spare
&
delimiter
The bits 9-16 determine the
PTU parameters included in
the composite data message
obtained with the following
commands:
-ASCII: aR0 and ar0
-NMEA 0183: aR0, ar0
-SDI-12: aM, aMC, aC, and aCC
-SDI-12 continuous: aR and
aRC
9th bit
Pa Air pressure
10th bit
Ta Air temperature
11th bit
Tp Internal temperature1
12th bit
Ua Air humidity
13th bit
spare
14th bit
spare
15th bit
spare
16th bit
spare
[I]
=
Update interval: 1 ... 3600 seconds
[P]
=
Pressure unit: H = hPa, P = Pascal, B = bar, M =
mmHg, I = inHg
[T]
=
Temperature unit: C = Celsius, F = Fahrenheit
=
Response terminator