Vaisala WXT510 User Manual
Page 54

User’s Guide ______________________________________________________________________
52 __________________________________________________________________ M210470EN-D
Example (ASCII and NMEA 0183, device address 0):
Changing the device address from 0 to 1:
0XU,A=1
1XU,A=1
Checking the changed settings:
1XU
1XU,A=1,M=P,T=1,C=2,I=0,B=19200,D=8,P=N,S=1,L=25,
N=WXT510,V=1.00
Example (ASCII, device address 0):
Changing RS-232 serial interface with ASCII, polled communication
protocol and baud settings 19200, 8, N, 1 to RS-485 serial interface with
ASCII, automatic protocol and baud settings 9600, 8, N, 1.
Checking the actual settings:
0XU
0XU,A=0,M=P,C=2,I=0,B=19200,D=8,P=N,S=1,L=25,N=WXT510,
V=1.00
Changing several settings with one command:
0XU,M=A,C=3,B=9600
0XU,M=A,C=3,B=9600
Checking the changed settings:
0XU
0XU,A=0,M=A,T=1,C=3,I=0,B=9600,D=8,P=N,S=1,L=25,
N=WXT510,V=1.00
NOTE
You can change several parameters in the same command as long as
the command length does not exceed 32 characters (including
command terminator characters ! or
those setting fields that are not to be changed.