Overview of MI70 Support
GMP231 is compatible with instruments that utilize the MI70 Indicator,
for example the GM70 Hand-Held Carbon Dioxide Meter. The MI70
indicator is a convenient service tool for viewing the measurement
readings, adjusting the environmental compensation settings, and
performing calibration and one-point adjustment.
1403-120
Figure 21
CO
2
Reading on MI70 Screen
Connecting GMP231 to MI70 Indicator
1.
If the GMP231 is installed in an incubator, disconnect the
incubator’s cable from the 8-pin M12 connector. Note that
GMP231 can remain physically connected to the incubator.
2.
If the MI70 indicator is on, turn it off.
3.
Connect the GMP231 to the MI70 indicator using the GMP231
service cable for MI70. See section Spare Parts and Accessories on
page 87.
4.
Turn on the MI70 indicator. MI70 detects the probe and proceeds
to show the measurement screen. The parameters measured by
GMP231 will start to show valid measurement results after a few
seconds.