beautypg.com

Example: set compensation temperature, Table 28, Example set_parameter invoke message – Vaisala GMP231 User Manual

Page 64: Table 29, Example set_parameter response message

background image

User's Guide _______________________________________________________________________

62 ___________________________________________________________________ M211501EN-C

Example: Set Compensation
Temperature

Table 28

Example Set_Parameter Invoke Message

Message Bytes (hex) Content
12

I

2

C address

82

Set_parameter command

09

Device address

0A

Frame length

10

Parameter ID of compensation temperature

00

Temperature 37C

00
14
42
0C

CRC high

6E

CRC lo


Table 29

Example Set_Parameter Response Message

Message Bytes (hex) Content
13

I

2

C address

04

Status byte (error active, ACK)

82

Set_parameter command

09

Device address

08

Frame length

10

Parameter ID of compensation temperature

00

Return code: Ok

D1

CRC high

CE

CRC lo