Figure 4, Measurement inside the incubator – Vaisala GMP231 User Manual

Page 18

background image

User's Guide _______________________________________________________________________

16 ___________________________________________________________________ M211501EN-C

1403-161

Figure 4

CO

2

Measurement Inside the Incubator

1 = Chamber wall.
2 = Chamber interior.
3 = Light source.
4 = Band pass filter.
5 = Ambient air (400 ... 1000 ppm CO

2

).

6 = Light absorbed by CO

2

in the incubator gas.

7 = Thermopile detector.
8 = Fabry-Perot interferometer.
9 = Sapphire window. Separates the sensor’s active components

from the measured environment.

10 = Gold-plated mirror.