beautypg.com

Set measurement to standby mode – Vaisala GMP231 User Manual

Page 39

background image

Chapter 4 ____________________________________________________ Vaisala Industrial Protocol

VAISALA ________________________________________________________________________ 37

Example below sets temperature compensation to setpoint mode, and
changes temperature setpoint value to 37.2 in RAM.

pass 1300
tcmode on
T COMP MODE : ON
env xtemp 37.2
In eeprom:
Temperature (C) : 37.00
Pressure (hPa) : 1013.20
Oxygen (%O2) : 19.70
Humidity (%RH) : 93.00

In use:
Temperature (C) : 37.2
Pressure (hPa) : 1013.20
Oxygen (%O2) : 19.70
Humidity (%RH) : 93.00

NOTE

If temperature and pressure compensations are configured to use
internally measured values, they will continuously update the values in
RAM, overriding any temperature and pressure values that are written to
RAM with the ENV command.

Set Measurement to Standby Mode

Use the STANDBY command to set the probe to standby mode. In
standby mode, the CO

2

measurement is disabled. Probe returns to the

normal measurement when the command STANDBY OFF is given or
the probe is reset.

STANDBY [on | off]

Example (enable standby mode):

standby on
Standby mode : ON