Stop measurement output, Set output interval, Output a reading once – Vaisala GMP231 User Manual
Page 34

User's Guide _______________________________________________________________________
32 ___________________________________________________________________ M211501EN-C
Stop Measurement Output
You can stop the measurement output by pressing the ESC key or with
the S command.
Set Output Interval
Use the INTV command to change the output interval of the
automatically repeating measurement messages. The measurement
messages are repeated in the RUN mode, or after the R command has
been given.
INTV [n xxx]
where
n
=
time interval, range 0 ... 255
xxx
=
time unit = "S", "MIN", or "H"
The shortest output interval (with n = 0) outputs the measurement
messages as quickly as the probe’s internal measurement cycle produces
them, without additional delay.
Example:
intv 5 s
Output interval : 5 S
Output a Reading Once
Use the SEND command to output a single measurement message.
SEND<cr>
Example:
send
CO2= 1702 ppm