Stop Measurement Output

You can stop the measurement output by pressing the ESC key or with
the S command.

Set Output Interval

Use the INTV command to change the output interval of the
automatically repeating measurement messages. The measurement
messages are repeated in the RUN mode, or after the R command has
been given.

INTV [n xxx]

where

n

=

time interval, range 0 ... 255

xxx

=

time unit = "S", "MIN", or "H"


The shortest output interval (with n = 0) outputs the measurement
messages as quickly as the probe’s internal measurement cycle produces
them, without additional delay.

Example:

intv 5 s
Output interval : 5 S

Output a Reading Once

Use the SEND command to output a single measurement message.

SEND<cr>

Example:

send
CO2= 1702 ppm