beautypg.com

Adjusting measurement, One point adjustment, Table 38 – Vaisala GMP231 User Manual

Page 67: Adjustment return codes

background image

Chapter 5 _______________________________________________________________ I2C Interface

VAISALA ________________________________________________________________________ 65

Table 38

Adjustment Return Codes

Code Meaning
0

Ok

1

Function not supported

2

Sequence error

3

Recorded-Reference difference too large

4

2-point adjustment: Points too close

5

Given reference value is invalid

Adjusting Measurement

NOTE

Before implementing an adjustment sequence using the Adjust command,
read the general instructions for calibration and adjustment in section
Calibration and Adjustment on page 73.


Create the reference environment inside the chamber, and use the Adjust
command to perform the adjustment sequence.

Adjustment can fail for a number of reasons. Refer to the list of return
codes for the Adjust command in Table 38 above.

One Point Adjustment

To perform a one point adjustment of carbon dioxide (CO

2

),

measurement using I

2

C protocol, you must perform the following

sequence:

1.

Using the Adjust command, start 1 point adjustment for the
chosen parameter.

2.

Place the sensor in the reference environment and wait until the
measurement has stabilized.

3.

Using the Adjust command, record the measured point and give
the reference value.

4.

Use the Adjust command to end the adjustment.