User's Guide _______________________________________________________________________

76 ___________________________________________________________________ M211501EN-C

I

2

C Interface

I

2

C interface supports one and two-point adjustment sequences as well as

configuration of the environmental compensation settings using the
Get_Parameter and Set_Parameter commands. See the following
sections:

- Adjust on page 64

- Get_Parameter on page 59

- Set_Parameter on page 61

Vaisala Industrial Protocol

Vaisala industrial protocol supports one and two-point adjustment with
the CCO2 command. You can also adjust the internal pressure and
temperature measurements with a simple offset correction. Configuration
of the environmental compensation settings can be done using serial line
commands. See the following sections:

- Calibration Commands on page 41

- Changing Measurement Settings on page 35

MI70 Hand-Held Indicator

MI70 hand-held indicator supports one-point adjustment, either using a
calibration gas or using a reference instrument that is connected to the
MI70. See section Calibration and Adjustment with MI70 Indicator on
page 79.