Configuring serial line operation, Set serial line operating mode, Table 11 – Vaisala GMP231 User Manual

Page 40: Selection of output modes

background image

User's Guide _______________________________________________________________________

38 ___________________________________________________________________ M211501EN-C

Configuring Serial Line Operation

Set Serial Line Operating Mode

Use the SMODE command to set the start-up operating mode of the
serial line.

SMODE [mode]

where

mode =

STOP, RUN, or POLL.


Table 11

Selection of Output Modes

Mode

Measurement Output

Available Commands

STOP

Only with the SEND command. All (default mode).

RUN

Automatic output.

Only command S.

POLL

Only with the SEND [addr]
command.

SEND [addr] and OPEN [addr].
Other commands available after
opening a line to the probe using
the OPEN command.

Use with RS-485 buses where
multiple probes can share the
same line.


Selected output mode will be activated at next reset or power up.

Example:

smode poll
Serial mode : POLL