User's Guide _______________________________________________________________________

42 ___________________________________________________________________ M211501EN-C

Example (show current user adjustment status – no adjustment done):

pass 1300
cco2
1.Ref. point low 0
1.Meas. point low 0
2.Ref. point low 200000
2.Meas. point low 200000
Gain : 1.0000
Offset : 0.0000

Example (one-point adjustment in 5% concentration (50000 ppm)):

pass 1300
cco2 –lo 50000
OK
cco2 –save
OK

Set Calibration Date

Use the CDATE command to view the date when the probe was last
calibrated, or set a new date.

CDATE [yyyy mm dd]

where

yyyy = Year of calibration in four digits
mm

= Month of calibration in two digits

dd

= Day of calibration in two digits


Example (show currently stored calibration date):

>pass 1300
>cdate
Calibration date : 2013-10-03

Example (set calibration date to 2014-03-10):

>pass 1300
>cdate 2014 03 10
Calibration date : 2014-03-10