Boonton 52000 Series USB CW Power Sensor User Manual
Page 63
![background image](/manuals/401801/63/background.png)
Boonton 52000 Series
52012
Frequency
10MHz to 12.4GHz
Measurement range
-50 to +20dBm CW
Operating Temperature
0 to 50 deg C
Shock
25G, 11 ms
Vibration
15G, 100 to 2000Hz
Linearity error 25 +/- 5 deg C
+/-0.27 dB; -40dBm to +10dBm;
greater than or equal to 50 MHz
+/-0.49 dB; +10dBm to +20dBm;
greater than or equal to 50 MHz
+/-0.28 dB; -40dBm to +10dBm; less than 50 MHz
+/-0.95 dB; +10dBm to +20dBm; less than 50 MHz
Cal factor error 25 +/- 5 deg C
+/-0.24 dB; 10 MHz to 50 MHz
+/-0.19 dB; 50MHz to 4.5GHz
+/-0.23 dB; 4.5GHz to 8.5GHz
+/-0.32 dB; 8.5GHz to 12.4GHz
Linearity variation 25 +/- 25 deg C
+/-0.29 dB; 50MHz to12.4GHz
+/-0.47 dB; 10 MHz to 50MHz
Cal factor variation 25 +/- 25 deg C
+/-0.29 dB; greater than or equal to 50 MHz
+/-0.44 dB; less than 50 MHz
Zero set
+/-1.2 nW
Noise
0.12 nW RMS
Input SWR - max
1.26:1 = 12.4 GHz 1.26:1 = 18.5 GHz
Measurement Speed: -50 to -35 dBm
8 measurements per sec.
-35 to -20 dBm
33 measurements per sec.
-20 to +20 dBm
50 measurements per sec.
Max input power (damage level)
200mW CW (+23dBm)
Connector type
SMA (m)
Connectivity
USB 2.0 (cable length up to 5m)
USB power supply current
Approx. 50mA
Sensor Cable Length (std)
76” (193 cm)
Dimensions (max) H x W x L
1.34” x 1.69” x 4.92” (34 x 43 x 125 (mm)
Weight
0.18lb (83 grams)
Specifications include expanded uncertainty of measurement stated as the standard uncertainty of
measurement multiplied by the coverage factor k=2 which corresponds to a coverage probability of
approximately 95% for abnormal distribution.