beautypg.com

2 using the software, Making a measurement – Boonton 52000 Series USB CW Power Sensor User Manual

Page 32

background image

2

Using the Software

8

1

2

3

4

5

6

7

8

1

TB

3

Data

57

57

Data

Training

26

f

1

f

1

TB

3

GB

8.25

1 TDMA frame (4.615 ms)

1 timeslot (0.577 ms)

TB = Tail Bits
f = Flag
GB = Guard Band

Figure 2 GSM TDMA structure and normal burst

In some cases, when the modulation is GPRS, multiple time slots are
used. The correct duty cycle can be found by multiplying 11.6% by
the number of active time-slots.

Making a measurement

Follow this procedure:

1. Start the application.

2. zero the sensor.
Refer to “Zero your sensor” on page 26.

3. Open the measurement control panel. Select the control panel icon
or select Measurement > Control Panel.

4. Enable the duty cycle option and set the duty cycle percentage to
11.6%.

5. Select Apply.
The sensor will read the GSM phone power in the active part of the
time slot with reasonable accuracy.