1.0 Introduction

1.1 Equipment Provider Responsibility

All Equipment Providers of the NewLife

®

Elite/NewLife

®

Oxygen Concentrator

must assume responsibilities for handling, operational check-out, patient

instruction, and maintenance. These responsibilities are outlined below and

throughout this manual.

WARNING

NewLife units must not be used for or with any life-

supporting applications. Geriatric, pediatric, or any other

patients unable to communicate discomfort while using

this machine may re quire additional monitoring. Advise

patients to immediately notify their Equipment Providers

and/or physicians in case of an alarm or any discomfort.

As an Equipment Provider, you must do all of the following:

Inspect the condition of each NewLife unit immediately upon

delivery to your business location. Note any sign of damage, external

or internal, on the delivery receipt, and report it directly to both the

freight company and AirSep Corporation immediately.

Check the operation of each NewLife before delivery to a patient.

Always operate the unit for a reasonable length of time and check that

the oxygen concentration level is within specifications as referred to in

Section 2.4. (Test the battery and power disconnect alarm as described

in Section 2.3 of this manual.)

Deliver NewLife units only to patients authorized by a physician’s

prescription. The NewLife must not be used as a life-supporting

device. A backup supply of oxygen must be available.

Instruct patients how to use the NewLife in conjunction with the

Patient Manual.

Instruct patients to notify their physicians and/or Equipment

Providers if they experience any signs of discomfort.

Instruct each patient how to perform routine maintenance of the air

intake gross particle filter and how to check the alarm system battery.

(Refer to Section 3.2.)

Be available to service each patient at any time.

Maintain the NewLife in accordance with Section 4.0.

Repair components and replace parts only as outlined in this

manual. Use only AirSep parts for replacement in NewLife Oxygen

Concentrators.

1-1