beautypg.com

1 compressor replacement – AirSep MN105-1 C User Manual

Page 22

background image

AirSep

®

Corporation

NewLife

®

Elite Service Manual

5.3.1 Compressor Replacement

Compressor Assembly Removal

To remove the compressor assembly for exchange, follow the steps

listed below:

1

Set the unit’s ON/OFF switch to the OFF position, and unplug

the power cord.

2

Remove the side and lower front panels.

3

Disconnect the suction tube.

4

Disconnect the blue compressor lead at the terminal strip and

the brown lead at the temperature switch.

5

Disconnect the two leads to the capacitor and remove if

necessary.

6

Disconnect the compression fitting for the heat exchanger

located at the bottom center of the compressor.

7

Remove the two screws that connect the compressor plate to

the base of the unit, and lift up and slide out the compressor

assem bly.

8

If needed, remove compressor from the plate by removing the

four compressor bolts.

Compressor Assembly Installation

To install the rebuilt compressor assembly, follow the steps listed below:

1

Perform the compressor removal procedure in reverse or der.

2

Make sure to position the compressor’s lead wires behind the

braided suction tube.

3

Leak test all connections.

5-3