NOTE

If a pressure transducer circuit board,

cut tie-wrap at circuit board and

disconnect green 1/4-inch tube. Plug/

cap open end of tube until new circuit

board is replaced.

4

Push in on the locking tab to the alarm 6-pin connector to

disconnect it.

5

With a slotted screwdriver, push in on the board support tabs,

while you lift each corner of the circuit board.

6

Remove the circuit board.

NOTE

Handle the new circuit board only

by the edges to prevent electro static

damage to the unit.

5.7.2 Circuit Board Installation

NOTE

If a pressure transducer circuit board,

first reinstall 1/4-inch green tube with

new tie-wrap.

1

Push the circuit board on to the support tabs.

2

Firmly plug in the 6-pin connector to the circuit board so that

the connector’s locking tab locks against the circuit board.

3

Plug in the 10-pin connector.

4

Reconnect the back and side panels.