beautypg.com

12 buzzer replacement, 13 hour meter replacement, 2 on/off power switch installation – AirSep MN105-1 C User Manual

Page 34

background image

AirSep

®

Corporation

NewLife

®

Elite Service Manual

5.11.2 ON/OFF Power Switch Installation

Follow the removal procedure for the ON/OFF power switch in reverse

order to install a new power switch. Make sure to reinstall the switch

properly by having the “O” on the switch located on the bottom when

finished.

5.12 Buzzer Replacement

1

Set the unit’s ON/OFF switch to the OFF position, and unplug the

power cord.

2

Remove the left side panel.

3

Unscrew the two Phillips-head screws that hold the buzzer to the

superstructure, and solder to disconnect the buzzer leads.

4

Solder the leads to the new buzzer, and screw the buzzer to the

superstructure.

5

Reconnect the side panel.

5.13 Hour Meter Replacement

1

Set the unit’s ON/OFF switch to the OFF position, and unplug the

power cord.

2

Remove the left side panel.

3

Disconnect the hour meter leads.

4

Push hour meter mounting tabs to remove the hour meter through the

control panel.

5

Install the new hour meter into the control panel, making sure that the

“hourglass” icon displays on the bottom.

6

Reconnect the hour meter leads.

7

Reconnect the side panel.

8

Plug unit in and set the unit’s ON/OFF switch to ON. Hourglass icon

should flash showing it is functioning correctly.

5-15