4 product regulator cleaning or rebuilding – AirSep MN105-1 C User Manual

Page 31

background image

AirSep

®

Corporation

NewLife

®

Elite Service Manual

2

Set the ON/OFF switch to the ON position, and allow the unit

to run at least three minutes to build up pressure.

3

Connect the humidifier bottle, oxygen tubing, and cannula to

the oxygen outlet.

4

Remove the right side panel.

5

Turn the flowmeter adjustment knob counterclockwise un til

it stops (wide open).

6

Pull outward on the regulator knob to unlock it.

7

Turn the regulator knob clockwise until the flowmeter ball

centers on the 5.5 lpm line.

8

Push in the regulator knob to lock the setting.

9

Turn the flowmeter adjustment knob clockwise until the

flowmeter indicates 5 lpm.

10

Reconnect the side panel.

NOTE

When this back pressure correction at

5 lpm is no longer required, fol low the

normal regulator setting procedure as

described in Section 5.8.1.

5.8.4 Product Regulator Cleaning or Rebuilding

Clean or rebuild the product regulator if the flowmeter ball fluc tuates

more than 1/4 of a liter or if the regulator cannot be adjusted for lockout.

1

Set the unit’s ON/OFF switch to the OFF position, and unplug

the power cord.

2

Remove the right side panel.

3

Use large pliers to unscrew the bonnet of the product regulator,

which contains a large spring.

5-12