beautypg.com

0 troubleshooting, 1 operating pressure test, 1 high operating pressure – AirSep MN105-1 C User Manual

Page 37

background image

AirSep

®

Corporation

NewLife

®

Elite Service Manual

6.0 Troubleshooting

6.1 Operating Pressure Test

Testing the operating pressure is a useful diagnostic tool when a concentrator

has low oxygen concentration and requires servicing. Units functioning

normally do not require operating tests.

Use the following procedure to test the operating pressure of the unit.

1

Set the unit’s ON/OFF switch to the OFF position, and unplug the

power cord.

2

Remove the right side panel.

3

Connect the pressure test gauge to appropriate test port location. (Some

models may require an additional test adapter (KI257-1) which allows

connecting directly to the 1/4-inch black pressure tubing from the valve

block.)

4

Plug in the power cord, and set the unit’s ON/OFF power switch to

the ON position. Set the flowmeter to 5 lpm, and allow the unit to run

at least five minutes.

5

Observe the maximum and minimum readings on the pressure test gauge.

The maximum reading should not exceed 32 psig (220.6 kPa). The

minimum reading should not be less than 10 psig (69 kPa).

NOTE

When you turn the unit on, the system pressure

always registers higher than normal for the first

few minutes of operation. Make sure the unit is

not in economy mode of operation if equipped.

6.1.1 High Operating Pressure

Higher than normal operating pressure may indicate any of the

following:

A restrictive blowdown muffler, which does not allow the

waste (purge) gas to exit the system freely.

Operate the unit with the blowdown muffler disconnected to

see if the operating pressure returns to normal.

6-1