beautypg.com

3 alarm system, 1 battery test, 2 power failure alarm test – AirSep MN105-1 C User Manual

Page 12

background image

AirSep

®

Corporation

NewLife

®

Elite Service Manual

d. OPTIONAL: The Oxygen Monitor’s amber light

remains on until the oxygen concentration reaches 85%

± 3% (approximately two minutes).

3

Turn the flowmeter adjustment knob counterclockwise until it stops

(wide open).

The maximum flow should register 5.5 lpm (the 5.5 lpm line should

appear to split the middle of the flowmeter ball) with no connections

at the oxygen outlet. If not, refer to Section 5.8.1 to set the product

regulator.

4

Perform an oxygen concentration test, as described in Section 2.4.

2.3 Alarm System

The NewLife Oxygen Concentrator is equipped with a battery-powered alarm

system, which sounds a continuous and loud alarm when a power failure

occurs. It sounds an intermittent alarm if the high or low pressure indicators

are activated or if the optional oxygen monitor detects lower than therapeutic

levels of oxygen concentration. The alarm remains on until you correct the alarm

condition or you set the ON/OFF switch to the OFF position. Refer to Section

6.0 for a list of probable alarm causes.

2.3.1 Battery Test

Each time the NewLife unit is turned on, a five-second audible alarm

sounds to indicate the condition of the battery.

NOTE

The audio alarm must sound loudly for

approximately five seconds each time the unit is

turned to ON to indicate the bat tery is in good

condition.

2.3.2 Power Failure Alarm Test

To test the power failure alarm, take the following step:

Unplug the unit, and set the ON/ OFF switch to the ON

position.

This should immediately activate the audio alarm. If it does not, refer

to the troubleshooting chart in Section 6.0 of this manual.

2-2