AirSep

®

Corporation

NewLife

®

Elite Service Manual

6-4

Problem

Probable Cause

Solution

Compressor does not run.

Constant audio alarm with

ON/OFF power switch in ON

position.

No power to unit.

Unit circuit breaker tripped or

faulty.

Faulty electrical connections.

Defective circuit board.

Defective ON/OFF power switch.

Check wall outlet for power.

Reset or replace circuit breaker.

Check electrical connections.

Check circuit board for two-light

rotation/replace circuit board.

Replace ON/OFF power switch.

Compressor runs with

intermittent alarm. Oxygen

concentration at 5 LPM is within

specifications.

Defective high or low pressure

switch.

Replace and retest pressure

switch.

Compressor shuts down

intermittently.

Restricted air flow through unit.

Unit overheating due to improper

location.

Defective cabinet fan.

Defective high temperature

switch.

Clean inlet filter, or remove

obstruction.

Locate unit away from heating

source, providing adequate

ventilation on all sides.

Replace cabinet fan.

Replace high temperature switch.

Compressor does not start.

ON/OFF power switch in ON

position, intermittent alarm, and

cabinet fan turns.

Extreme cold start.

Compressor thermally cut off due

to excessive heat. NOTE: It will

not restart unit until it cools down.

Defective capacitor.

Defective circuit board.

Faulty electrical connection for

compressor.

Defective high temperature

switch.

Allow unit to reach room

temperature.

Blocked air intake or defective

cabinet fan/clear obstruction, or

replace cabinet fan.

Replace capacitor.

Check circuit board for two-light

rotation/replace circuit board.

Check electrical connections for

compressor.

Replace high temperature switch.

6.3 Troubleshooting Chart