AirSep

®

Corporation

NewLife

®

Elite Service Manual

An improperly operating circuit board or solenoid valve.

Confirm that the circuit board and solenoid valves function

properly.

Contaminated sieve beds.

Change the sieve beds.

6.1.2 Low Operating Pressure

Lower than normal operating pressure may indicate any of the

following:

A restriction in the suction resonator or optional air intake

filter, which limits the amount of room air available to the

compressor.

Disconnect the braided suction tube at the compressor, and

allow the unit to operate without the suction resonator to see if

normal operating pressure returns.

An improperly operating circuit board or solenoid valve.

Confirm that the circuit board and solenoid valves function

properly.

A leak in the unit, which allows system pressure to escape.

Leak test the unit.

A compressor with reduced output.

Ensure that the oxygen concentration level at the desired

liter flow is within AirSep’s specifications. If it is below

specifications, replace or repair the compressor.

An obstructed or vertically positioned (ECOnomy mode)

EcoCheck valve.

For the EcoCheck valve (available only as an option for units

equipped with an oxygen monitor): confirm that the EcoCheck

valve is not obstructed or in the ECOnomy mode.

6.2 General Troubleshooting

Before reviewing the troubleshooting chart, the following steps may be useful

to isolate any malfunctions:

1

Turn the concentrator on. If unit does not turn on, refer to

troubleshooting chart.

6-2