beautypg.com

2 feed or waste valve reassembly, 3 solenoid valve coil replacement – AirSep MN105-1 C User Manual

Page 24

background image

AirSep

®

Corporation

NewLife

®

Elite Service Manual

5

Loosen and remove the solenoid base with a one-inch deep

well socket.

6

Install the rebuild kit, which contains all parts of the sole noid

assembly except the solenoid coil.

5.4.2 Feed or Waste Valve Reassembly

To reassemble the feed or waste valves, follow the feed or waste valve

rebuilding procedure in reverse order, and test for leaks.

5.4.3 Solenoid Valve Coil Replacement

If a solenoid valve coil does not operate, its corresponding green light

on the circuit board does not illuminate.

1

Set the ON/OFF switch to the OFF position, and unplug the

power cord.

2

Remove the side and back panels.

3

Remove the retaining cap with a slotted-head screwdriver.

4

Disconnect the solenoid leads, and lift off the solenoid coil.

5

Replace with the new coil.

6

Reinstall the retaining cap back on top of the coil, and reconnect

the solenoid leads.

7

Reconnect the back and side panels.

5-5