beautypg.com

AirSep MN105-1 C User Manual

Page 36

background image

AirSep

®

Corporation

NewLife

®

Elite Service Manual

10

Pull the power cord through the opening on the power cord receptacle.

11

Using same distance located on old power cord, slide the strain relief

onto the new power cord with the lip of the strain relief facing the ends

of the power cord leads.

12

Insert the new power cord through the opening on the receptacle by

squeezing the strain relief together with pliers.

13

Position the strain relief upward, and push it into the rear of the power

cord opening on the receptacle.

14

Reconnect the power cord leads, close the plastic twist clamps, and

reattach the tie-wraps.

15

Reconnect the power cord receptacle to the base of the unit with the

two Phillips-head screws.

16

Reconnect the side and back panels.

5-17