Elite Service Manual

4.2.1 Preparing for New Patient Use

When you remove the NewLife from a patient’s home, always dispose

of the used nasal cannula and humidifier bottle. Clean the exterior of

the NewLife with a soapy water solution or commercial cleaner to

remove any gross debris, organic or otherwise. Be careful not to get

any liquid into the interior of the unit.

Next, clean the exterior with either a common chemical disinfec tant

or a bleach solution* before other patients use the unit. For the bleach

solution, wear eye and skin protection to prevent exposure to the

chlorine. Allow the NewLife to air dry, and then retest it before you

return it to inventory.

Clean the air inlet gross particle filter with warm soapy water between

each patient’s use. Clean this filter at least once per week, depending

on the environment, during normal operation.

Change the product filter after 25,000 hours of use. It is not necessary

to change this filter between patients even if the pre vious patient had

a communicable disease or infection.

________________________________

*Make the bleach solution a 1:100 dilution of 5.25% sodium hypochlorite. Mix one part household bleach (e.g.,

Clorox) with 99 parts cold tap water. To measure the solution easily, take 1/4 cup of household bleach, and mix it

with a gallon of cold tap water. Allow the mixture to sit on potentially contaminated surfaces for 10 minutes.