10 circuit breaker replacement, 11 on/off power switch replacement, 1 circuit breaker removal – AirSep MN105-1 C User Manual
Page 33: 2 circuit breaker installation, 1 on/off power switch removal

AirSep
®
Corporation
NewLife
®
Elite Service Manual
5
Install the new pressure switch, connect the leads, and leak test.
6
Reconnect the side panel.
5.10 Circuit Breaker Replacement
5.10.1 Circuit Breaker Removal
1
Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2
Remove the left side panel.
3
Disconnect the circuit breaker leads.
4
Unscrew the circuit breaker while you apply pressure to the
circuit breaker retaining ring.
5.10.2 Circuit Breaker Installation
Follow the removal procedure for the circuit breaker in reverse order
to install the new circuit breaker.
5.11 ON/OFF Power Switch Replacement
5.11.1 ON/OFF Power Switch Removal
1
Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2
Remove both side panels, the lower front panel, and the back
cover.
3
Unscrew the four Phillips-head screws that hold the control
panel to the superstructure.
4
Cut off the two white tie-wraps from the terminal strip leads.
5
Disconnect the ON/OFF switch leads from the back of the
switch.
6
Push on the back of the power switch, while holding in its
retaining tabs, and remove the switch through the front panel.
5-14