“Specifications continued”

Medical equipment needs special precautions regarding EMC and need to be
installed and put into service according to the EMC information provided in this
section.

Guidance and manufacturer’s declaration ± electromagnetic immunity

The Focus is intended for use in the electromagnetic environment specified below. The customer or the user of the
Focus should assure that it is used in such an environment.

IMMUNITY test

IEC 60601

test level

Compliance level

Electromagnetic environment

±

guidance

Electrostatic
discharge (ESD)

IEC 61000-4-2

± 6 kV contact


± 8 kV air

± 6 kV contact


± 8 kV air

Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.

Electrical fast
transient/burst

IEC 61000-4-4

± 2 kV for power
supply lines

± 1 kV for
input/output lines

± 2 kV for power
supply lines

Not Applicable

Mains power quality should be that of a typical
commercial or hospital environment.

Surge

IEC 61000-4-5

± 1 kV line to line

± 2 kV line to earth

± 1 kV line to line

± 2 kV line to earth

Mains power quality should be that of a typical
commercial or hospital environment.

Voltage dips, short
interruptions and
voltage variations on
power.

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5s

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the Focus
requires continued operation during power
mains interruptions, it is recommended
that the Focus be powered from an
uninterruptible power supply (UPS) or
a battery.

Power frequency
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.