Focus™ en – AirSep MN172-1 User Manual
Page 217

Focus™
EN
AirSep
®
Corporation
EN-29
MN172-1 rev. A - 02/13
How to Respond to Focus’s Audible Alarm and
Indicator Lights
Status
Audible Alarm
Light
Indicates
Action
Indicator Brief, continuous at
start–up
(Green) and
(Red) alternate;
then (Green)
continuous light
Focus has been
turned on.
You may begin to
operate your Focus
unit.
Indicator No
(Green)
flashes;on each
breath
Focus is delivering
oxygen as a pulse
flow.
Continue using Focus
normally.
Indicator No
(Amber)
continuous light
Inspection and/or
Service required.
Return unit to
Equipment Provider
for inspection and/or
service.
Battery
Indicator
No
25% (Green)
light; flashes
Battery charge is
low.
Connect the Focus
unit into a DC outlet
or an AC outlet
immediately. Charge
battery.
Alarm
Intermittent:
Beep
(Amber) alarm;
intermittent light
Warning: Battery
voltage
approaching too
low a level to
continue operating
Focus.
Connect the Focus
unit into a DC outlet
or an AC outlet
immediately.
Charge battery.
Alarm
Intermittent:
Beep, beep
(Amber) alarm;
intermittent light
Battery shutdown:
Battery voltage is
too low to operate
Focus.
Connect the Focus
unit into a DC outlet
or an AC outlet
immediately.
Charge battery.
Alarm
Continuous:
Beep
(Amber) alarm;
continuous light
No breath detected
by the unit for a
predetermined
time period.
Check the cannula
connection. Ensure
that you are breathing
through your nose.
If the alarm persists,
contact your
Equipment Provider.