MN172-1 rev. A - 02/13

How to Respond to Focus’s Audible Alarm and
Indicator Lights

Status

Audible Alarm

Light

Indicates

Action

Indicator Brief, continuous at

start–up

(Green) and
(Red) alternate;
then (Green)
continuous light

Focus has been
turned on.

You may begin to
operate your Focus
unit.

Indicator No

(Green)
flashes;on each
breath

Focus is delivering
oxygen as a pulse
flow.

Continue using Focus
normally.

Indicator No

(Amber)
continuous light

Inspection and/or
Service required.

Return unit to
Equipment Provider
for inspection and/or
service.

Battery

Indicator

No

25% (Green)
light; flashes

Battery charge is
low.

Connect the Focus
unit into a DC outlet
or an AC outlet
immediately. Charge
battery.

Alarm

Intermittent:
Beep

(Amber) alarm;
intermittent light

Warning: Battery
voltage
approaching too
low a level to
continue operating
Focus.

Connect the Focus
unit into a DC outlet
or an AC outlet
immediately.
Charge battery.

Alarm

Intermittent:
Beep, beep

(Amber) alarm;
intermittent light

Battery shutdown:
Battery voltage is
too low to operate
Focus.

Connect the Focus
unit into a DC outlet
or an AC outlet
immediately.
Charge battery.

Alarm

Continuous:
Beep

(Amber) alarm;
continuous light

No breath detected
by the unit for a
predetermined
time period.

Check the cannula
connection. Ensure
that you are breathing
through your nose.
If the alarm persists,
contact your
Equipment Provider.